Detection Method: Infrared Controlled Interference
Polarization Spectrometer
Measurement Range: 0 to 10,000 PPM; 1 to 100% Gas
Display Range: Auto Ranging PPM: 0-10,000 %Gas: 1-100%
Manual Ranging PPM scale: 0-10,000 %Gas scale: 0-100%
Sensitivity: 1 PPM
Resolution: 0-1000 PPM: 1 PPM
1000-10,000 PPM: 5 PPM
1-100%Gas: 0.1%
Accuracy: Greater of ±0.5% or ±10% of reading (normal temperature and pressure,
% gas on manual mode)
Response Time:* PPM Mode: T10< 0.7s, T50< 1.3s
PPM Mode with Survey Probe: T10< 1.2s, T50< 1.8s
%Gas Mode with Method 21 Probe: T50< 1.8s, T90< 2.5s;
Recovery Time T10< 2.0s
Sample Flow Rate:* 2.0 LPM (with Survey Probe attached)
Connectivity:* Bluetooth 5 Low Energy (BLE) (compatible with iOS, Android and
Windows devices)
Detection Alarms
Modes: Digital Methane Detection (DMD):
Audible tone when detection threshold exceeded. Adjustable Alarm
Level from 1 to 9,000 PPM
Tick: Continuous audio tone relative to concentration
System Fault Warning: Audible tone and indication on the display
Self Test & Calibration: Built in Self Test and Calibration function verifies operation and
adjusts calibration for maximum sensitivity. Test gas cell
integrated within the instrument.
Compliance: EN 61326-1
Conducted Emissions Class B EN 61326-1
Radiated Emissions Class B EN 61000-4-2 4kV/8kV
ANSI C63.4 Class B EN 61000-4-3 3 V/M
FCC 47 CFR Part 15 Class B
Dust Filter: Replaceable filter provides dust protection. Easy change-out quick
disconnect filter cap.
Display: Large easy to read backlit LCD (0.75” Numeric)
Operating Temperature: 0 to + 122 F (-17 to 50 C) (nominal battery voltage)
Humidity: 5 to 95% RH, non-condensing
Enclosure: IP54 (Water splash and Dust resistant)
Instrument Weight: 5.6 lbs.
Carry Case: 13 lbs empty; 21 lbs filled 24.5” x 21” x 9”
Power Supply: Internal rechargeable Li-ion battery or External 12 Vdc car battery
with power conditioner
Battery Operating Life: 8 hours at 32º F without backlight on
Battery Charger: External, in-line, 110 - 240 VAC, 50 / 60 hertz. 10 hours to 90% charge
Shoulder Strap: Single over the shoulder padded strap
Alarm Volume: 108 dBs @ Alarm port (A-fast)
Survey Probe: Quick connect fitting with locking clasp. Adjustable length
from 25” to 41”
Bar hole Probe: Std: 20”; Optional 36”
Method 21 Probe (optional): Std: 5’, with quick connect fitting and locking clasp
Intrinsic Safety: Class I Division I Group D T3
UL 913, CSA 22.2 No. 157, MET Labs #E112840
* These features only available on DP-IR+ released January 2020 onwards.
|